CALIBrAtIOn PrOCeDUre

Do not use inert gases to zero the monitor.  This will cause premature failure of the sensor.

CO/O2 Monitor Zero Adjustment

 

To zero the monitor, follow the steps below.  Zero calibration gas should be used to properly “zero” the monitor 

and assure that a valid calibration is achieved.  If zero adjustment cannot be made as indicated, sensor replacement may 

be necessary.  After each monitor adjustment outlined in the steps, allow time for the changes to stabilize.

SteP 1) 

 

Place the “ON/OFF/TEST” switch in the “ON” position.

SteP 8) 

 

Allow digital readout to stabilize approximately 15-30 

seconds.

SteP 10)

  

Turn the regulator off and disconnect the regulatorfrom 

the zero gas cylinder.

SteP 2)

Allow 30 seconds for the readout to stabilize.  The green 

indicator will illuminate.

SteP 3)

Hold the “ON/OFF/TEST” switch in the “TEST” position.  

The following will occur:

 

Audible alarm will sound

 

“LOW O2” and “NORMAL” indicators blink

             together and alternate with the “HIGH CO”

             indicator.

 

Amber “LOW BATTERY” indicator on monitor

             will illuminate.

This test ensures the circuitry is operable and continuity 

to the sensor is proper. Release the switch.

SteP 4)

Remove the air sample inlet tube.

SteP 5)

Install regulator on the zero air cylinder reference gas.

SteP 7)

Attach the clear tubing with the male plug to the air sam-

ple inlet on the monitor.

SteP 6)

Turn the knob on the regulator counterclockwise to allow 

the fl ow of gas thru the hose. Verify fl ow of gas thru the 

hose via touch or sound.

SteP 9)

Adjust the “zero” adjustment screw (clockwise to increase 

or counterclockwise to decrease) until a reading of “00” is 

obtained.

Calibration Procedure

Содержание CO2-91IS

Страница 1: ...ne 757 424 3967 Toll Free 1 800 866 8100 Fax No 757 424 5348 www airsystems com e mail sales airsystems com This manual must be read carefully and followed by all persons who have or will have the res...

Страница 2: ...Ct Or InDIreCt COStS eXPenSeS LOSSeS Or DAMAGeS ArISInG OUt OF DeFeCtS In Or FAILUre OF tHe PrODUCt Or Any PArt tHereOF The purchaser shall be solely responsible for compliance with all applicable Fed...

Страница 3: ...nitor will analyze the air sample and display the oxygen and CO concentration in parts per million ppm The system s green NORMAL operation light will illuminate and the red HIGH CO light will icker ap...

Страница 4: ...tor lights and audible alarm mounted on bottom of housing and Model CO2 91ISLA has the indicator lights mounted on top of housing with the audible alarm being mounted on the left endplate Carbon Monox...

Страница 5: ...CO 91BEP 19 Audible Alarm CO2 91ISLA ELLS004IS 20 Aluminum Housing MONC021A 21 Led Socket MONC009LA 22 Yellow LED MONC008NS 23 LED Socket And Yellow LED CO 91LED 24 Red LED CO2 91ISLA MONC035NS 25 Gr...

Страница 6: ...teP 3 Remove endplate to gain access to the sensor cups SteP 4 Remove sensor from sensor cup and remove leads Take the new sensor and remove the metal clip CO sensor only Reattach leads to the proper...

Страница 7: ...conds for the readout to stabilize The green indicator will illuminate SteP 3 Hold the ON OFF TEST switch in the TEST position The following will occur Audible alarm will sound LOW O2 and NORMAL indic...

Страница 8: ...nterclockwise to decrease until the digital readout reads the same as the concentration ppm as printed on the calibration gas cylinder SteP 6 Turn the regulator off and repeat the zero adjustment proc...

Страница 9: ...9 Notes...

Страница 10: ...10 Notes...

Страница 11: ...11 Notes...

Страница 12: ...sapeake Va 23320 Telephone 757 424 3967 Toll Free 1 800 866 8100 Fax No 757 424 5348 http www airsystems com e mail sales airsystems com Printed in the U S A Copyright Air Systems International Inc 20...

Отзывы: