CHAPTER I 
 
BASIC PRODUCT INFORMATION 

Based on our vertical laminar flow cabinet product line, our PCR cabinet has been designed specifically 
for polymerase chain reaction (PCR) applications. The cabinet isolates critical steps in the PCR process 
by providing a particulate-free work environment and can b

e operated effectively as “mini-environments” 

in the laboratory. Vertical laminar flow cabinet offers proven protection for your sample and processes 
where operator protection is not required. They are suitable for application which requires ISO Class 3 air 
cleanliness within work zone as per ISO 14644-1 (equivalent to Class 1 per US Federal Standard 209E) 

 
AIRFLOW PATTERN INSIDE THE CABINET 

Room air is taken in from the top of the cabinet through a disposable pre-filter with 85% arrestance; this 
serves to trap larger particles and increase the life of the main filter. Air is forced evenly across an 
HEPA/ULPA-filtered vertical laminar flow air stream within the cabinet; the result is a stream of clean 
laminar air within the work zone of the cabinet; this dilutes and flushes all airborne contaminants from the 
interior.  A nominal filter face velocity of 0.45 m/s or 90fpm ensures that there is sufficient number of air 
changes within the enclosed are of the cabinet in order to maintain cleanliness. The purified air travels 
across the internal work zone of the cabinet in a vertical, unidirectional stream and leaves the main work 
chamber across the entire open front of the cabinet. 

  
PURAIR-PCR Airflow Pattern    

 

 

 

Содержание PURAIR-PCR

Страница 1: ...AIR SCIENCE PURAIR PCR CABINETS USER SERVICE MANUAL Air Science Manual Revision No PURAIR PCR SERIES V3 2014 Specifications subject to change without notice...

Страница 2: ...of performance validation certification 11 3 4 2 Re certification 11 3 4 3 Disclaimer 11 IV Operating Your Cabinet 12 4 1 Laminar Flow Control System 12 4 2 Cabinet Start Up Procedure 12 4 3 Working...

Страница 3: ...tand the installation procedures and take note of the environmental electrical requirements of the cabinet In this manual important safety related points will be marked with this symbol If the equipme...

Страница 4: ...on the use of Hazardous Substances RoHS With respect to the directive on RoHS please note that this cabinet falls under category 9 monitoring and control instruments and is therefore exempted from req...

Страница 5: ...e evidence of loss or damage may result in the carrier refusing to honor a claim Contact delivering terminal to arrange for a claim form and inspection report to be faxed or mailed to you Notify Air S...

Страница 6: ...limited warranty does not cover Installation inside delivery handling damage Products with missing or defaced serial numbers Consumables such as filters HEPA ULPA carbon pre filters and fluorescent UV...

Страница 7: ...120 6 th Street Fort Myers Florida 33907 USA www Airscience com Tel 239 489 0024 Fax 239 489 0922 6 USER SECTION...

Страница 8: ...quivalent to Class 1 per US Federal Standard 209E AIRFLOW PATTERN INSIDE THE CABINET Room air is taken in from the top of the cabinet through a disposable pre filter with 85 arrestance this serves to...

Страница 9: ...he cabinet is protected by Styrofoam cardboard and or and shrink wrap If you find any damage during this stage of unpacking please refer to freight claim information t is the best practice to leave th...

Страница 10: ...ny related equipment Some modifications to the work surface may be necessary 2 2 PACKAGING CONTENTS The following items are included together with your manual Test certificate Test report In case this...

Страница 11: ...rature range from 20 deg C to 30 deg C 68 deg F to 86 deg F It is recommended that the temperature in the laboratory be maintained within 2 deg C under all conditions 5 Main supply voltage fluctuation...

Страница 12: ...he cabinet as it is heavy 3 4 PERFORMANCE VALIDATION CERTIFICATION After installation and prior to use cabinet performance must be validated and certified to factory standards The following tests shou...

Страница 13: ...ep the back air grilles unobstructed by arms or objects Always surface decontaminate everything before inserting them into the work zone 4 Load and arrange the materials apparatus to minimize the move...

Страница 14: ...r flow cabinets is explicitly discouraged in all major international standards and recommendations 2 Once complete turn off the cabinet remove laboratory coat and gloves and wash hands using germicida...

Страница 15: ...ing the interior the operator should not insert any part of his her body into the work zone except hands Monthly 1 Using a damp cloth clean the exterior surfaces of the cabinet particularly the front...

Страница 16: ...ENANCE SCHEDULE Model Year Serial Number Responsible Person Month Clean exterior surface Check all service fixtures By Who Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec Daily Surface Decontamination...

Страница 17: ...120 6 th Street Fort Myers Florida 33907 USA www Airscience com Tel 239 489 0024 Fax 239 489 0922 16 SERVICE SECTION...

Страница 18: ...rgo recertification after filter changing Before the new filters are installed all surfaces should be thoroughly cleaned of silicon and or adherent gasket material The new filter should be carefully h...

Страница 19: ...TER FOR DAMAGE EXAMINE FRAME FOR BYPASS SEAL WITH SILICONE VERIFY THAT LEAK IS NOT A CABINET LEAK RE EXAMINE FRAME ROTATE FILTER END FOR END 180 DEGREES REPLACE FILTER AND RE TEST IF FRAME LEAKS IF FR...

Страница 20: ...With leveling feet 2 IV Bar with 6 hooks For IV bags filling applications 3 Service Fixtures both European and American style fixtures available To provide supplies of gas vacuum water and compressed...

Страница 21: ...r faulty relay section If the motor does not run the motor is defective Replace motor Inoperative Switches Connection Problem Check whether the cable from electrical panel has been connected properly...

Страница 22: ...bags and put it into the outside trash or other protected outside location for the next normal trash collection Never send a fluorescent light bulb or any other mercury containing product to an incin...

Страница 23: ...mments if any at the bottom and continue on another sheet if necessary We are also interested to receive comparison against any competitors if you are able to provide the information Excellent Very Go...

Страница 24: ...ights reserved The information contained in this manual and the accompanying product are copyrighted and all rights are reserved by Air Science Air Science reserves the right to make periodic minor de...

Страница 25: ...lation Information Date Cabinet Model Company Serial Number Responsible Person Please record any major maintenance procedures performed by the technician and or operator including filter changes recer...

Страница 26: ...t Fort Myers Florida 33907 USA www Airscience com Tel 239 489 0024 Fax 239 489 0922 25 For units equipped with this option please follow the following On line Screens for set up MENU Screen Set 1 MENU...

Страница 27: ...120 6 th Street Fort Myers Florida 33907 USA www Airscience com Tel 239 489 0024 Fax 239 489 0922 26 MONITAIR WIRE DIAGRAM...

Страница 28: ...120 6 th Street Fort Myers Florida 33907 USA www Airscience com Tel 239 489 0024 Fax 239 489 0922 27...

Отзывы: