O2 Detector-Calibration
GCT recommends users of the O2 Single Gas Clip
to bump test the detector before using each day.
Single Gas Clip Oxygen (O2) detector factory default will
prompt the user to calibrate the detector every 30 days. The
user will be prompted by the screen flashing CAL, please
see calibration instruction below:
Calibration Instructions
Only perform O2
calibration in normal air
(20.9% Oxygen) that is
free of hazardous
gases.
1.
Press and hold down the
yellow button for four (4)
seconds
2.
The screen will display CAL
and the O2 icon will flash in the
lower left hand side
3.
After a successful
calibration, the detector will
emit one (1) beep, vibration
and LED flash
4.
After an unsuccessful
calibration, the detector will
beep, flash, and continue to
display calibration. If after a
few failed calibration please
contact GCT customer support
at 877-525-0808
Cleaning
The detector can be cleaned with a soft damp cloth. Do not
use solvents, soaps or polishes. A neutral cleaner, like Mat &
Table Top Cleaner (by
ACL Staticide
) may also be used.
The sensor screen may be cleaned with a soft brush under
clean, warm water. Return the filter to the detector once it
has fully dried.
Hibernation (SGC+ Only)
When the SGC Plus is not used for extended period of time
it can be turned off (Hibernated) to suspend the 24 month
operation life countdown.
SGC Plus Hibernation with the IR Link
1. Check that you have installed the IR Link Software and the IR Link
USB connections are plugged in
2. Click Read Device on the IR Link Software
3. Note when the monitor is hibernated the event log will be cleared. It
is highly recommended to save the event log by pressing the save
event log before hibernating.
4. Click on the Hibernate Button, acknowledge the event log message
5. Keep in front of the IR Link until the "Hibernate OK" message is
displayed at the bottom of the IR Link Software
6. Confirm the monitor screen is blank
7. If you encounter any problems please contact GCT customer
support at 877-525-0808
SGC Plus Hibernation with the Clip Dock
1. Check that the Clip Dock is turned on and USB memory is inserted
2. The Clip Dock is capable of hibernating 4 units at one time, place
the desired amount of monitors in the docking station
3. Press and hold the bump test and calibration buttons down
simultaneously for approximately 2 seconds
4. A successful hibernation will result in a GREEN light on for the
corresponding unit number
5. Note the event log will be automatically stored on the Clip Dock
USB memory
6. Confirm the monitor screen is blank
7. If you encounter any problems please contact GCT customer
support at 877-525-0808
Single Gas Clip and SGC Plus User’s Manual Rev
2.05
Single Gas Clip and SGC Plus User’s Manual Rev
2.05