background image

- INSTRUCTIONS -

WIRING DIAGRAM:

NH-1SA-AC            NH-2SA-AC

Patient Stations

Single and Dual Patient Stations for All Call,

and when sharing a common Corridor Lamp

NHX-80X

11
12
13
14
15
16

21
22
23
24
25
26

~

51
52
53
54
55
56

A+

A-

B+

B-

NOTES:

1. All Call requires one of the station buttons to be used for All Call.
Therefore, the standard system can have up to 49 stations, and the expanded
system can have up to 79 stations.
2. Only wiring concerning the NH-1SA-AC and NH-2SA-AC is shown here.  For
complete installation and wiring information, please see the NHX Installation
instructions.

PS-2420UL

+

-

Trunkage Line

Connection

Terminals

Some applications for Nurse Call systems require All Call.  By including the NH-50AC adaptor and the "AC" model
patient stations, a station button on the master can be used to call all of the patient stations simultaneously.  An
additional wire is required to each patient station. In addition, if there is more than one patient station sharing a
corridor lamp, this model patient station will allow that configuration to be used.

NOTE: 

An NHX system installed with an NH-1SA-AC or NH-2SA-AC included does not carry the UL 1069 listing.

NHX

Nurse Center System

CN1:

Brn

Red

Org

Yel

Grn

Blu
Blk

CN2:

Brn

Red

Org

Yel

Grn

Blu
Vio

Two Patient Stations in one room sharing a bathroom and corridor lamp

NH-1SA-AC

CN1:

Brn

Red

Org

Yel

Grn

Blu
Blk

CN2:

Brn

Red

Org

Yel

Grn

Blu
Vio

NH-2SA-AC

Brn

Red

Org

NHR-4A

Yel

Grn

Blu

Bathroom Pullcord

(Wired to one station only)

PS-2420UL

+

-

Aiphone Communication Systems
1700 130th Ave. N.E.
Bellevue, WA  98005
(425) 455-0510
FAX (425) 455-0071

NH-1/2SA-AC  Instr.

0103js

TOLL FREE TECHNICAL SUPPORT:

(800) 692-0200

TOLL FREE FAX LINE:

(800) 832-3765

E-MAIL: [email protected]

APPLICATION #1:

All Call feature added, where an announcement can
be made from the master station to all patient
stations simultaneously.

APPLICATION #2:

2 patient stations in a room with a common
bathroom and one corridor lamp outside the room.

* Bath call will annunciate on
one position at the master
station, according to which
patient station it is wired to.

To NH-50AC

NHR-7A

Corridor Lamp

(Shared by both patient stations)

Отзывы: