4
Test Control:
Untimed, manually timed or automatically timed tests; user-defined test time.
Declaration of load class. Limits automatically determined from EN61000-3-2
for appropriate class; Class C and Class D limits can be automatically
calculated from power measurements or from ratings declared by the user.
Minimum and maximum power thresholds for limits can be changed by the
user.
Facility for declaring supply voltages other than 230V and deriving
appropriate limits. Facility for insetting test limits.
Report Printing:
Direct printer connection for hard-copy report with user-entered narrative,
supply voltage assessment and current harmonic analysis and assessment.
VOLTAGE FLUCTUATIONS & FLICKER METER
Measurements:
Voltage fluctuations d
c
, t
max
and d
max
and flicker P
st
and P
lt
to EN61000-3-3 and EN61000-4-15.
Current Rating:
16A rms continuous, or national connector rating if lower.
Voltage Ranges:
115V (± 200V pk) and 230V (± 400V pk).
Fluctuation Range:
25% max (relative to nominal voltage).
Flickermeter Range:
Voltage change up to 20% (sinewave change) or 10% (low repetition rate
rectangular change) relative to AGC level.
Equivalent to 6400 pu on 8.8 Hz sinewave.
Flickermeter AGC:
Up to ±5%.
Flickermeter Accuracy:
Better than 5% for P
st
range 0·7 to 10·0.
Frequency Range:
50Hz (operates over 43 – 67 Hz).
Report Printing:
Tabular listing of voltage variations, P
st
classifier and P
st
in each P
lt
interval.
GENERAL
Display:
320 x 240 pixel backlit LCD.
Clock:
Real
−
time clock for time and date
−
stamping of Report data.
Interfaces:
Parallel Printer, RS-232 (for printer or PC); also USB (for PC) on HA1600A.
Instrument Supply:
220V–240V or 110V–120V ±10%, 48 to 65Hz. Installation Category II.
Maximum instrument power 25VA. The instrument can be operated at a
different supply voltage and/or frequency from that used to supply the load.
Operating Range:
+5°C to 40°C, 20-80% RH.
Storage Range:
–10°C to +60°C.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Safety & EMC:
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this instrument via
http://www.aimtti.com/support (serial no. needed).
Dimensions (WxHxD):
HA1600A: 357 x 132 x 235mm;
HA1600: 305 x 148 x 220mm.
Weight:
HA1600A: 4.4kg;
HA1600: 4.2kg.