General

1

Examine all the warning signs on the bed, examine the bed for obvious signs of damage. 

All aspects of the equipment should operate as intended. Check that all nuts, bolts and 

other fasteners are tight and are not missing.

2

Examine flexible cables and conduits for cuts, cracks, abrasions or other deterioration.

3

Check that the power supply plug is not damaged. If either the power cable or plug is 

damaged then both the cable and plug must be replaced as a complete assembly by an 

approved service agent.

Central Locking System

1

Apply the brakes and push the bed forwards. If any of the four braking castors rotate the 

brake is not fully effective.

2

Check the two steering lock castors for correct operation.

3

If either the brakes or the steering locks are not working correctly contact an approved 

service agent.

Side Rails (if fitted)

1

•  The locking pin mechanism and striker plate arrangement should be checked every 

three months.

•  Inspect the side rails for general wear and tear on fittings and pivot joints every six 

months.

•  Check for rust, structural integrity of welds and paint finish every twelve months.

2

Raise the side rail to the upright position until the locking mechanism clicks into 

place. Refer to the operating procedures starting on page 10 of this manual for further 

instruction.

3

Check that the there is no yellow shaft visible. In the locked position the red on the 

locking pin shaft should not be visible from above

4

If the red shaft is visible when the side rail is in the raised and locked position, contact 

Aidacare for service advice.

5

Check that the rails pivot axles are securely fitted to the bed. The pivot axles are locked 

into place by the manufacturer by a 10mm lock nut. 

NOTE: Other than general maintenance activities all other service or repair 
work must be performed by qualified and trained persons who are approved 
by Aidacare.
WARNING: The bed must be disconnected from the mains power supply 
before starting any maintenance activity.

Содержание BEB045900

Страница 1: ...USER MANUAL AC SERIES BEDS Code Model BEB045900 Single 3 Function BEB045905 King Single 3 Function BEB045901 Single 2 Function...

Страница 2: ...au Thank you for choosing this Aidacare AC Series Bed This product has been specifically designed to offer comfort support and safety It is important that you read this user manual carefully before u...

Страница 3: ...1 Introduction 2 Safety Precautions 3 Electronic Functions 4 General Bed Operation 5 Specifications 6 Cleaning 7 Maintenance 8 Transportation and Storage 9 Accessories 10 Warranty and Service 4 5 7 9...

Страница 4: ...s and Area of Use The bed is appropriate for adult patient care under supervised medical monitoring Key Features Maximum safe working load SWL of the bed is 220kg Electrically adjusted height range of...

Страница 5: ...ughout this manual identify possible hazards in procedures or conditions which if not followed correctly could result in equipment damage or failure GENERAL Electrical equipment can be hazardous if us...

Страница 6: ...s when the bed is stationary ALWAYS store the handset when not in use to avoid accidental operation of the beds functions The handset can be stored by hooking it on to the side rail of the bed ALWAYS...

Страница 7: ...cal power cable from the mains and store before transporting the bed If the power cable shows any sign of damage replace immediately CAUTION Do not operate the bed controls for more than the prescribe...

Страница 8: ...gers are free from potential squeeze hazards Knee Positioning Adjustment To move the knee positioning section up and down press and hold the appropriate button on the handset The knee section is adjus...

Страница 9: ...eleased and the bed can be moved in any direction With the pedals in the raised Steer Track position one steering castor at the head end of the bed is engaged to provide better control when pushing th...

Страница 10: ...mately 10mm of thread length will allow for the internal lock nut to travel freely past the thread of the hand wheel 2 With both hand wheels now unscrewed gently pull with an even force on both sides...

Страница 11: ...when fitted correctly fold down towards the foot end of the bed 1 Before fitting the rails unscrew the hand wheel to bring the locking clamp into an open neutral position which will allow the rail to...

Страница 12: ...released WARNING Care should be taken whilst operating the side rails to ensure hands and fingers are kept away from potential squeeze hazards Side rails should only be used after clinical assessment...

Страница 13: ...included for use when a foot board is not required on the bed and should be removed when a foot board is in place To extend the mattress retainer remove the two retaining pins and pull out until the r...

Страница 14: ...r Input 230V AC 50Hz Max current 1A Rate Load Interruption 10 Max 2 minutes 18 minutes Leakage Protection Class 1 Type B Waterproof Grade IP54 Without Side Rails 112 Kg With Side Rails 132 Kg S W L Sa...

Страница 15: ...0Hz Max current 1A Rate Load Interruption 10 Max 2 minutes 18 minutes Leakage Protection Class 1 Type B Waterproof Grade IP54 Without Side Rails 117 Kg With Side Rails 137 Kg S W L Safety Factor 2 x S...

Страница 16: ...lean with NaDCC e g Presept Actichlor at 10 000 ppm of available chlorine Rinse and dry Allow the cleaned parts to dry before replacing the mattress CAUTION Do not use abrasive Hypocarbonate or Phenol...

Страница 17: ...te arrangement should be checked every three months Inspect the side rails for general wear and tear on fittings and pivot joints every six months Check for rust structural integrity of welds and pain...

Страница 18: ...sure 50 kPa To 106 kPa Ambient temperature 10 C To 40 C Relative humidity 30 To 75 Air pressure 70 kPa To 106 kPa Product code Description BEA029700 Folding Side Rails BEA029705 Self Help Pole Stainle...

Страница 19: ...and conforms to all ACCC guidelines around consumer rights For further information contact Aidacare on 1300 133 120 or email product aidacare com au Aidacare will not warrant the safety and or correct...

Страница 20: ...Imported by AIDACARE P L 4 Noonan Rd Ingleburn NSW 2565 T 1300 133 120 W www aidacare com au ABUM08 1 110815...

Отзывы: