Commissioning 

3. Activate the relay using the command ´

ipp f1 Rpp

´,

Deactivate the relay using the command ´

ipp f1 R-pp

´.

Application Driving the analog output :

1. Configure analog output port 6 or 7; see Manual 6.10.7:

e.g. selected analog type x (command ´

ipp f9 Ax

´)

assigning to the selected measuring point (command ´

ipp f9 E-00

´)

or assigning to any reference channel xx (command ´

ipp f9 Exx

´)

or output the value yyyyy via the interface (command ´

ipp f9 ayyyyy

´)

2.  Define the desired measuring range for the measuring point, to be output

over the whole analog output range (0 to 10 V or 0 to 20 mA), using the pa -

rameters analog start and analog end in sensor programming, special func-

tions, see Manual 6.10.7.

3. If for a current output the output range 4 to 20 mA is to be used (instead of 0

to 20 mA), the element flag for 4 to 20 mA must be activated for the appro-
priate measuring point; see Manual 6.10.3. 

8. Display and operating controls 

The adapter incorporates a graphics display that can show exactly all states
affecting the interface elements. This is configured largely by the device itself. 

8.1 Main menu 

The   main   menu   is   called   up   by   pressing

<MENU> 

. Here, by pressing 

 

▲ 

 / 

 ▼ 

, you can

access menu items 

 All ports 

 or 

 Single ports 

or 

 Device configuration  

and then, by press-

ing 

 ► 

 or 

 PROG 

, activate these. 

By means of  

  <☼ON> 

  you can switch on dis-

play illumination.

8.2 Showing all elements :

Below the port number the type of elements
available is shown with its abbreviation. Below
this the activation status of the relays and trig-
ger inputs is shown and in line xy23 the actual
switching status of the relays and the type of
analog outputs. 

8.3 Showing single elements :

You   can   by   pressing  

  ▲  

  and  

  ▼  

  scroll

through and select any port individually and
display details of address, element type, vari-
ant, and status.
With  

relays

  one   can   also   see   the   actual

switching status of the individual contacts.

 ALMEMO

®

 8006-RTA3

7  

Содержание ALMEMO 8006-RTA3

Страница 1: ...____________________________ Operating instructions Relay trigger adapter analog ALMEMO 8006 RTA3 V2 4 18 03 2014 www ahlborn com...

Страница 2: ...uts TR8 TR9 2 DC socket Mains adapter ZA1312 NA1 12V 0 2A 3 LCD graphics display 7 rows for functions 1 line for softkeys F1 F2 Shown in brackets MENU 4 Operating keys TR8 9 Trigger keys softkeys P Po...

Страница 3: ...4 Connecting peripheral equipment 5 6 COMMISSIONING 6 7 DISPLAY AND OPERATING CONTROLS 7 7 1 Main menu 7 7 2 Showing all elements 7 7 3 Showing single elements 7 7 4 Device configuration 8 7 5 Messag...

Страница 4: ...maximum supply current on the measuring instrument must take the sensors into consideration If the maximum supply current is exceeded a mains adapter e g ZA1312 NA1 should be connected at the DC socke...

Страница 5: ...with electrically iso lated analog outputs offering the following signals Option Output signal Gain OA 8000 R2 0 000 to 10 000 V 0 5 mV digit or 0 000 to 20 000 mA 1 A digit The output value normally...

Страница 6: ...uring points program limit values on the ALMEMO device 2 With ALMEMO device program the first relay port to variant 0 Summated alarm using command i20 f9 k0 see Manual 6 10 9 2 3 In the event of any l...

Страница 7: ...appro priate measuring point see Manual 6 10 3 8 Display and operating controls The adapter incorporates a graphics display that can show exactly all states affecting the interface elements This is c...

Страница 8: ...desired function can be selected by means of keys PROG and Programming is by means of the softkeys ON OFF or SET and input is by means of keys PROG PROG Illumination To switch display illumination ON...

Страница 9: ...Watchdog Duration s o Duration 060s Switch off watchdog alarm with watchdog function and switch on again after restoring the control 8 5 Messages Each relay can be assigned a particular message either...

Страница 10: ...output port pp ipp f9 ayyyyy Program message for port pp e g ipp message port pp Start new line with special character Tel 123 CRLF To output message Response ipp P48 Message port pp Tel 123 CRLF To...

Страница 11: ...ty 50 V 0 5 A Trigger inputs Optocoupler 4 to 30 V Input current 2 mA Analog outputs Electrically isolated switchable OA 8006 R02 4 00 to 10 0 V 0 5 mV digit Load 100 kW 0 0 to 20 0 mA 1 A digit Load...

Страница 12: ...mp connectors ZA 8006 RTA3 Mains adapter with ALMEMO connector 12 V 1 A ZA 1312 NA8 Options 2 additional relays maximum 10 including 1 ALMEMO clamp connector OA 8006 SH2 2 additional normally closed r...

Страница 13: ...nection diagram 2 2 contact assignment 8 3 8 Current consumption 11 1 11 Device configuration 8 4 8 Display and operating controls 8 7 display illumination 8 4 8 driven inversely 9 10 Driving the anal...

Страница 14: ...Power supply 5 4 Power supply 11 1 11 Product overview 11 2 12 Programming via the device interface 9 10 Relay 11 1 11 relays 8 3 7 screw connector 6 4 5 Showing all elements 8 2 7 Showing single elem...

Страница 15: ...Index trigger source 8 3 8 variant 8 3 7 voltage capacity 6 1 4 Watchdog 8 4 9 8 4 8 ALMEMO 8006 RTA3 15...

Страница 16: ...11 Appendix 11 4 your contact 16 ALMEMO 8006 RTA3...

Отзывы: