Appendix
15. APPENDIX
15.1 Technical data
Measuring inputs :
2490-1
1
ALMEMO
®
socket suitable
for all ALMEMO
®
sensors
2490-2
2 ALMEMO
®
sockets, electrically isolated, suitable
for ALMEMO
®
sensors
Measuring channels:
4 channels / socket for double sensors, function
channels
2490-2
4 internal additional channels
A/D converter :
Delta - sigma, 16-bit, 2.5 / 10 mops, adj. 1 to 100
Sensor power supply :
9 volts, max. 400 mA (with OA2450-U only 80 mA)
Outputs :
2 ALMEMO
®
sockets suitable for all output modules
RS485 interface, integrated:
OA 2490-I, electr. isol., integrated, socket DC
Signals :
RX+, RX-, TX+, TX-, line, maximum 1 km
Analog output, integrated:
electrically isolated, socket P0
Variants:
2490-1R02, 2490-2R02
Outputs, options:
0.00 V ...+10.0 V 0.5 mV/digit Load > 100kΩ
0.0 mA ...+20.0 mA 1 µA/digit Load < 500Ω
Accuracy:
± 0.1% of final value
Temperature drift:
10 ppm / K
Time constant:
100 us
Standard equipment :
LCD :
Measured value : 5x 7-segment 15 mm, 2x 16-seg-
ment 9 mm
Function 4½ x 7-segment 9 mm, 9 symbols
Operation :
7 silicone keys
Memory
99 measured values on the RAM
Date and time-of-day Software clock, buffered by battery supply
Power supply :
Batteries :
3 AA alkaline batteries
Current consumption :
approx. 16 mA (without input and output modules)
With double analog output:
approx. 90 mA + 3.5 x I
OUT
External :
ALMEMO
®
socket DC
Clamp connectors
ZA1000-FSV 10 to 30 VDC
Mains adapter :
ZA 1312-NA1 230 VAC to 12 VDC, 0.2 A
Adapter cable, electr. isolated
ZA 2690-UK 10 to 30 VDC to 12 VDC, 0.25 A
Option U, electr. isolated
OA2490-U 10 to 30 VDC, 0.1A
Housing :
(LxWxH) 127 x 83 x 42 mm
ABS (acrylonitrile butadiene styrene),
weight : approx. 260 g
Suitable conditions
Operating temperature :
-10 ... +50 °C (storage temperature : -20 ... +60 °C)
Ambient relative humidity :
10 to 90 % rH (non-condensing)
ALMEMO
®
2490
33
(see Manual 2.3)
Содержание ALMEMO 2490-1
Страница 39: ...Your contact ALMEMO 2490 39...
Страница 40: ...15 Appendix 40 ALMEMO 2490...