respiratory treatment facilities in hospital or clinics. The more
restrictive acceptance limits of Group 1 Class B (CISPR 11) have been
considered and applied. The equipment is suitable for use in the
mentioned environments when directly connected to the Public Mains
Network.
b) The test is not applicable in this environment unless the ME
EQUIPMENT and ME SYSTEM used will be connected to the PUBLIC
MAINS NETWORK and the power input is otherwise within the scope
of the Basic EMC standard.
Guidance and manufacturer’s declaration - Electromagnetic
immunity - Enclosure port
Phenomenon
Basic EMC
standard or
test method
Immunity test levels
Professional
healthcare
facility
environment
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2 ± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
Radiated RF EM IEC 61000-4-3
a)
10 V/m
b)
80MHz - 2.7
Содержание AGU NS12