AGS8830 Operations Manual
-III-
The AGS8830-8,AGS8830-16 Real-Time PCR Detection System
is comply with the Standard IEC61010-1
:
2010, and the protection
degree is IP20.
It is intended for indoor use and applied to an elevation below
2,000 meters.
It is an in vitro diagnostic (IVD) medical equipment which is
comply with the standard IEC 61010-2-101: 2015.
a) Instrument grounding
WARNING: To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
The mains plug or appliance coupler is intended to be used as disconnecting
device from mains supply, not to position the equipment so that it is difficult to
operate the disconnecting device.
Please use the equipment in a manner specified by the manufacturer, or the
protection provided by the equipment will be impaired.
The input power cord for the instrument must be grounded reliably so as to
avoid electric shock. The instrument is equipped with 3-pin plug with a grounding
pin, and only mated with a grounding power socket. If the plug cannot be inserted
into the socket, please ask an electrician to install the right socket but never make
the grounding plug lose its safety protection function.
Warning: Against replacing detachable MAINS supply cords by
inadequately RATED cords.
Warning: please use parts supplied only by the manufacturer or his
agent.
b) Keep away from the circuits when the power is on.
The operator shall never remove the protection device, replace a component,
or adjust the instrument. If it is required, the manufacture shall complete it and it
is forbidden to replace any component when the power is on.
c) Attention to the power supply
Before AC power is applied, make sure that the power voltage is in line with
the value required by the instrument (24VDC 47-63Hz 9.2A).