Aginova Sentinel RS-485 DO Скачать руководство пользователя страница 10

8

2

Click  “Connection”  on  the  menu  bar,  select  the  first  line  in  the  drop-down  menu  “Connection 

setup”

The baud rate for the first time is the slave label

  and click “OK”. 

Note:

 

Port is set according to the Port number of the connection. 

Note:  If  the  sensor  has  been  connected  as  described,  and  “Timeout  Error”  appears  on  the 

software “Display status”, it means that the connection is failed; remove and replace the USB port 

or  check  the  USB  to  RS485  converter,  repeat  the  above  procedure  until  the  sensor  connection  is 

successful. 

Chapter 5 Calibration of Sensor 

The dissolved oxygen sensor has been calibrated at the factory and if you need to calibrate yourself, 

follow the steps below   

The specific steps of automatic air calibration are as follows: 

Dry the sensor, add a small amount of water (25~50mL) to the calibration kit and shake the bag for a 

few  times. 

Then  put  the  sensor  into  the  calibration  kit  without  touching  the  water  and  seal  the 

calibration kit with your hand to make the sensor in a saturated air.

 The sensor should be away from 

light and high temperature and hard objects during calibration. The measured data is stable, Double-click 

the “06”, and a dialog box pops out. Enter “27” for Address in the dialog box, “16” for Value,

 

then click 

“Send”. 

Tel

+1 513 204 5837

 

Web

http

s

://www.

aginova.

com

Sentinel (RS-485) DO •

 Operation Manual

Содержание Sentinel RS-485 DO

Страница 1: ...Sentinel RS 485 DO Operation Manual...

Страница 2: ...side fixed installation 5 3 1 3 Railing fixed installation 6 3 2 Connection of Sensor 7 Chapter 4 Interface and Operation 7 4 1 User Interface 7 4 2 Parameter Setting 7 Chapter 5 Calibration of Senso...

Страница 3: ...aterials SUS316L Ordinary Version Titanium Alloy Seawater Version Up and down cover PPS glass fiber Cable PUR Power Supply 9 28VDC Communication Protocol MODBUS RS485 Storage Temperature 15 60 Operati...

Страница 4: ...ely used in the DO online monitoring of the regulating tank biochemical tank and effluent of sewage treatment plant DO online monitoring in waterworks surface water water used in various industrial pr...

Страница 5: ...ion diagram of Sentinel RS 485 Dissolved Oxygen Sensor Chapter 3 Installation 3 1 Installation of Sensors Note The protective cap should be removed before use before measurement Do not screw down the...

Страница 6: ...4 3 1 1 Quick Dismantling pool side fixed installation Figure 3 Quick Dismantling pool side installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Страница 7: ...5 3 1 2 Classic pool side fixed installation Figure 4 Classic pool side fixed installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Страница 8: ...6 3 1 3 Railing fixed installation Figure 5 Railing fixed installation sketch map Tel 1 513 204 5837 Web https www aginova com Sentinel RS 485 DO Operation Manual...

Страница 9: ...onnect Note Modbus Poll software is a general software that can be downloaded online 4 2 Parameter Setting 1 Click Setup on the menu bar select Read Write Definition and then set the parameters The sl...

Страница 10: ...ed oxygen sensor has been calibrated at the factory and if you need to calibrate yourself follow the steps below The specific steps of automatic air calibration are as follows Dry the sensor add a sma...

Страница 11: ...ensor and place it in the anaerobic water until the measured data is stable Double click the 06 and a dialog box pops out Enter 27 for Address in the dialog box 12 for Value then click Send Wait 20 se...

Страница 12: ...n the following table MODBUS RTU Baud Rate 4800 9600 19200 38400 57600 Data Bits 8 bit Parity Check no Stop Bit 1bit Register Name Address Data Type Register Number Read Write Description Dissolved Ox...

Страница 13: ...60s multiples of 2 Air Calibration Step 1 27 Signed 1 W 16 Step 2 27 Signed 1 W 19 It should be cancelled if you don t want to calibrate after the execution of Step 1 Cancel 27 Signed 1 W 21 Zero poi...

Страница 14: ...n code 03 for reading register content 04 The length of the returned data is 4 bytes 00 00 40 E0 The slope read is 7 00 analyze 40 E0 00 00 using IEEE 754 CA 7B CRC16 check code 2 Modify Slave Address...

Страница 15: ...equipment return 01 10 00 06 00 02 A1 C9 Send command parsing 01 device address 01 10 Function code 16 for writing register content 00 06 The starting register address write is 00 06 00 02 Write 2 reg...

Страница 16: ...equently leading to the wrong reading 7 2 Inspection on the Damage of Sensor According to the appearance of sensor to check if there is damage if any damage is found please contact after sales service...

Страница 17: ...d when the cap is found severely damaged during the inspection Steps to replace the fluorescent cap Unscrew the old cap from the sensor and then screw the new cap on Note When removing and installing...

Отзывы: