17 Technical Information
Window Meanings
244/288
Turbo-V 1KG Rack Controller User Manual / 87-901-045-01
Soft Start
“Soft Start” mode is provided to start the pump after a protracted
stop of the Turbo-V 1KG Pump. This allows a better grease
distribution in the bearings.
The “Soft Start” mode is disabled by default. The “Soft Start” mode
can be activated by a suitable software, by Remote I/O or by Front
Panel command (see the paragraphs “INTERCONNECTIONS” and “RS
232/485 COMMUNICATION DESCRIPTION”). “Soft Start” frequency
steps are the followings: 300 Hz, 350 Hz, 400 Hz, 500 Hz, 700 Hz and
761 Hz. The pump remains at each step for a period of 300 s. A fail
condition occurs if the pump is not able to properly spin up the
driving frequency (Too High Load Error).