User’s and Service Guide U3047-90001
43
U3047AM12
Control Lines
External Voltage Supply Configuration
illustrates an example of the connection with an external dc power supply. Connect the positive
and negative voltage supply from the external power supply to the positive input (pin 13) and the negative
input (pin 14). Connect the power supply dc ground to pin 15.
Turning On the Test Set using an External Power Supply
1. Turn
on
the test set.
2. Connect the DUT.
3. Turn
on
the external power supply.
Turning Off the Test Set using an External Power Supply
1. Turn
off
the Power Supply.
2. Turn
off
the test set.
3. Disconnect the DUT.
Figure 37
Test Set to the DUT and External DC Power Supply
8
1
15
14
13
TO DUT (Only required Lines)
GND
+
Gnd
Gnd
DC Power Supply
12
13
14
15
1
2
8
2 to 5 Vdc
TEST SET
Line 1
Line 2
Line 8
GND
Positive: 0 to 5 Vdc
Negative: 5 to 0 Vdc
NC
9
(Rear Panel View)