QUICK REFERENCE GUIDE 

AGILENT TWISTORR 304FS 

4

3. Controller Connection and Start up Sequence 

ELECTRONIC CONNECTION 

A Power supply (Controller) is necessary to drive and operate the Turbopump.  Connect the controller cable to the pump connector. 

If turbopump has been stored or not used for a long period of time (> 6 months), the soft start procedure might be necessary. To 

enable the soft start refers to the dedicated «How to enable the soft start» paragraph.

SWITCH ON THE TURBO PUMP

1) Via RS232  - Tplus software

●  Connect the controller to the pump

●  Connect a RS232 cable from a PC to the controller Serial 

Port.

●  Power on the controller

●  Launch theTPlus Software. Enable the Serial mode.

●  Connect the 15 pin mating connector to J1 connector 

●  Start the primary backing pump and reach a vacuum 

pressure of  < 1 mbar

●  Pump can be started by pressing the «Start» button.

●  To stop the pump select the Stop button and decelerate 

the rotor by opening the vent screw of about ¼ to ½ turn.

●  To vent the pump see the «Vent procedure» paragraph

2) Plug and pump:

●  Connect the power supply

●  Connect the pump cable to controller.

●  Start the primary backing pump and reach a vacuum 

pressure of  < 1 mbar

●  Connecting the 15 pin mating connector to J1 connector 

the pump will start immediately. 

●  To stop the pump remove the J1 connector and 

decelerate the rotor by opening the vent screw of about 

¼ to ½ turn.

304 ON BOARD NAVIGATOR CONTROLLER 

There are 2 ways to start the pump: using the 

Tplus Software or by «plug and Pump»

304 AG RACK CONTROLLER 

●  Plug the 9 pin interlock connector (provided together 

the controller) in the P1 IN connector. If not «waiting 

interlock» message will appear and pump will not start).  

Connect the main power cord.

●  Start the primary pump and reach a vacuum pressure of  

< 1 mbar

●  Configure the controller to operate from the Front Panel 

by pressing simultaneously the two upper buttons to 

enter into the configuration menu and scroll till the 

«configuration is displayed: select FRONT.

●  Press again the two upper button simultaneously to exit 

from configuration menu.

●  Press the START button to run the pump. Press the 

STOP button to stop the pump.

●  To vent the pump please see the «vent Procedure» 

paragraph.

 

Содержание TwisTorr 304 FS

Страница 1: ...tailed reference guide to install and operate the TwisTorr 304 FS 1 Accessories and Configuration 2 Preparation for installation 3 Controller Connection and Start up sequence 4 Purge Vent valve 5 How...

Страница 2: ...ary Vane Technology Aluminium Block Valves Controller Unit Suggested Agilent Foreline pumps SH 110 IDP 15 TS 300 DS 42 DS102 202 302 Active Gauges Vent Valve Water Cooling Air Cooling Purge Valve On b...

Страница 3: ...100 160 4xM10 Claw Clamps with fixing torque 22Nm CFF6 16 screws with fixing Torque 20Nm CFF 8 20 screws with fixing torque 20Nm For water cooling flow of 50 l hr is mandatory if ambient temperature i...

Страница 4: ...Vent procedure paragraph 2 Plug and pump Connect the power supply Connect the pump cable to controller Start the primary backing pump and reach a vacuum pressure of 1 mbar Connecting the 15 pin mating...

Страница 5: ...tings 304 AG RACK CONTROLLER PURGE VALVE There are several models of purge valve available with different inlet flanges and orifices 10 or 20 SCCM to protect the bearings from particulates corrosive g...

Страница 6: ...e when powered on To operate the pump via SERIAL and activate the soft start click on Remote button If a PC is not available the soft start can be still be activated by using the 15 pin mating connect...

Страница 7: ...system Verify vent and purge screws valves to be closed or properly working Check the gas load in the chamber Check if rotor spins freely if not replace the pump PUMP OVERTEMPERATURE Navigator LED is...

Страница 8: ...or code 64 Rack AG message I displayed Loss of insulation motor windings in relation to ground Ramp currents too high Vbus control not correct One or more Mosfet is are faulty shorted after start or i...

Отзывы: