Operational and assay guidelines for non-37 °C
assays
•
For all non-37 °C operation, the XFe Analyzer must equilibrate overnight in
the required ambient temperature.
•
If it is required to set up the XFe Analyzer in a cold room, avoid direct fan
sources.
•
For all non-37 °C operation, the tray heater must remain
ON
. Do
NOT
turn the
tray heater
OFF
.
•
For assay temperatures below 30 °C, hydrate the Sensor Cartridge in the dark
at room temperature.
•
Prior to starting an assay, an additional 30 minutes of precalibration
equilibration time has been added to ensure temperature stability.
To adjust the Target Temperature (set point) using the up/down arrows, do the
following:
1
Click the Temperature widget. (See
The Tray Heater dialog box is displayed. (See
Figure 11.
Temperature widget
Figure 12.
Edit temperature settings (Tray Heater dialog box)