7
BenchCel Microplate Handling Workstation User Guide Addendum
7. Calculate the jog distance:
Jog distance = Z
initial
–
Z
adjusted
In the following example, the jog distance is –5.00 mm.
8. Calculate the adjusted Orientation sensor offset:
Adjusted Orientation sensor offset =
Initial Orientation sensor Jog distance
In the following example, the adjusted orientation sensor offset is
3.00 mm.
9. Type the adjusted orientation sensor offset in the
Labware
tab.
10. Select the next stack, and then click
Move to Sensor.
Check the Notch
Sensor readings. The notches should have much lower readings than
the corners. In addition, the difference between the notch readings
and corner readings should be at a maximum.
11. Repeat step 10 for the remaining stacks.
Z
initial
50.27 mm
Z
adjusted
55.27 mm
Jog distance
–5.00 mm
Initial Orientation sensor offset
8.00 mm
Jog distance
–5.00 mm
Adjusted Orientation sensor offset
3.00 mm
Содержание R Series
Страница 4: ...Letter to our Customers 2 Agilent Technologies Inc Automation Solutions Division ...
Страница 10: ......
Страница 23: ......
Страница 24: ......
Страница 26: ......
Страница 34: ...Preface BenchCel Microplate Handling Workstation User Guide viii ...
Страница 60: ...Chapter 1 Introduction BenchCel Microplate Handling Workstation R Series User Guide 26 ...
Страница 124: ...Chapter 4 Performing a run BenchCel Microplate Handling Workstation R Series User Guide 90 ...
Страница 188: ...Appendix A BenchCel ActiveX control BenchCel Microplate Handling Workstation R Series User Guide 154 ...
Страница 210: ...Appendix B Quick reference BenchCel User Guide 176 ...
Страница 218: ...Index BenchCel Microplate Handling Workstation R Series User Guide 184 ...
Страница 219: ......
Страница 220: ...Agilent Technologies User Guide G5400 90001 ...