Running Samples
84
NovoCyte Quanteon Flow Cytometer Operator's Guide
Compensation Channel
, select
FITC
,
PE
,
PerCP
, and
APC
. Select
Compensation
on Height
and select
Median
as
Parameter for Calculation
(Figure 50). Click the
OK
button and a new
Compensation Specimen
will appear in the
Experiment
Manager
under the experiment file node. This
Compensation Specimen
contains
a series of single-stained samples as selected in the
Compensation Channels
Figure 49.
New Auto Compensation Menu
Figure 50.
New Auto Compensation Window