Chapter 2
57
Test Descriptions
Noise Figure Range and Accuracy
Test Setup
Figure 2-14
Equipment required for Noise Figure Range and Accuracy test
Test Procedure
NOTE
Throughout the Noise Figure Range and Accuracy Test Procedure the term ‘the
Worksheet’ refers to the Noise Figure Range and Accuracy Test Worksheet on page
60 and the term ‘the Test Record’ refers to the Noise Figure Range and Accuracy
Test Record detailed in the relevant Appendix for the model number being tested.
Step 1.
Connect the equipment as shown in Figure 2-14.
Step 2.
Set the Synthesized Sweeper as follows:
1. Press the
Preset
key.
2. Set the frequency to 50 MHz (press the
CW
key and
5
,
0
,
MHz
.
3. Set the power level to -54 dBm (press the
Power Level
key and
-
,
5
,
4
,
dB(m)
.
4. Switch
RF ON
Step 3.
Set the Switch Driver as follows:
NOISE FIGURE
ANALYZER
ATTENUATOR
SWITCH DRIVER
SYNTHESIZED
SWEEPER
Control
Cable
Attenuator X
RF Cable
Adapter 2
Input
RF Cable
Adapter 1
PRECISION 1dB
STEP ATTENUATOR
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...