Chapter 2
15
Test Descriptions
10 MHz Out Frequency Reference Accuracy
Required Test Equipment
The frequency standard provides the reference for the Universal Counter. You can
use a company-specific standard in place of the 5071A, providing it meets the
critical specifications as listed in Table 2-1.
Test Setup
Before starting the 10 MHz Out Frequency Reference Accuracy Test, connect the
equipment as shown in Figure 2-1.
Figure 2-1
10 MHz Out Frequency Reference Accuracy Test Setup
Table 2-1
10 MHz Out Frequency Reference Accuracy Test Equipment
Equipment description
Critical specification for equipment
substitution
Recommended model
Universal Counter
Time Interval Range: 25ms to 100ms
Single Operation Range: +2.5 to -2.5Vdc
53132A
Frequency Standard
Frequency: 10 MHz Timebase Accuracy
(Aging): < 1 x 10
-10
/day
5071A
BNC Lead (X 2)
Length 122cm (48 in.), Frequency
10 MHz
10503A
NOISE FIGURE
ANALYZER
FREQUENCY
STANDARD
UNIVERSAL
COUNTER
BNC Cable
10 MHz
Ref Out
Ref In
10 MHz
BNC Cable
Channel 1
Содержание NFA Series
Страница 25: ...13 2 Test Descriptions This chapter descibes each test required for calibration and performance verification...
Страница 114: ...104 Chapter3 Technical Specifications General Specifications...
Страница 126: ...116 AppendixA Model N8973A Test Records Measurement Jitter Test Record...
Страница 138: ...128 AppendixB Model N8974A Test Records Measurement Jitter Test Record...
Страница 152: ...142 AppendixC Model N8975A Test Records Measurement Jitter Test Record...
Страница 178: ...168 AppendixD Caring for Connectors Principles of Microwave Connector Care...
Страница 184: ...174 Index Index...