Dynamic Contact Module II B
Agilent Nano Indenter G200 User’s Guide
B-14
8
To install the new tip in the DCM II head reverse the above steps.
9
When you have reached the point of screwing the new tip into the
head, slowly rotate the micro-socket wrench clockwise
(
Ensure a snug fit;
DO NOT
over-tighten.
Figure B-20
Turn wrench clockwise to screw in new tip
10
Remove the bottom-most locking pin.
11
Remove the top-most locking pin.
If the tip is not calibrated, follow the “Determining the Indenter Area
Function – DCM II” procedure.
If the tip has been calibrated, then:
1
Select
Select Tip
from the Tip menu, as shown in
.
Figure B-21
Select Tip command on Tip menu
Содержание Nano Indenter G200
Страница 1: ...Agilent Technologies Agilent Technologies Nano Indenter G200 User s Guide...
Страница 14: ...Agilent Nano Indenter G200 User s Guide 9 Express Test Tip Calibration H 12 Express Test Varied Force H 12...
Страница 96: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 4 Export Menu Mode Menu System Menu Help Menu...
Страница 109: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 17 Figure 4 18Load Review page configuration dialog box...
Страница 117: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 25 Figure 4 27Real Time Graph Setup dialog box...
Страница 121: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 29 Figure 4 31User Configuration dialog box...
Страница 136: ...NanoSuite 4 Agilent Nano Indenter G200 User s Guide 4 44 Exit Use this icon to exit the software...
Страница 263: ...Commonly Used Procedures 6 Agilent Nano Indenter G200 User s Guide 6 44 Figure 6 62Flowchart 1 Figure 6 63Flowchart 2...
Страница 478: ...Agilent Technologies Agilent Nano Indenter G200 User s Guide Part Number G2A 13192 1 Rev C Agilent Technologies Inc 2012...