3-36
Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
N5242A
10 MHz Frequency Reference Adjustment
The 10 MHz frequency adjustment is used to adjust the frequency accuracy of the network
analyzer’s 10 MHz frequency reference on the A10 frequency reference board assembly.
Equipment Used for the 10 MHz Frequency Reference Adjustment
Procedure
NOTE
This adjustment typically adjusts to within
±
0.01 ppm.
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the frequency counter.
Figure 3-13 Equipment Setup for the 10 MHz Frequency Reference Adjustment
2. Press UTILITY
, then
, then
, then click
10 MHz Freq Adjustment
.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
Equipment Type
Model or Part Number
Alternate Model or Part Number
Cable, BNC, 50
Ω
, 24 inch
8120-1839
Any
Frequency counter
53151A, Option 001
Any that will measure a signal at 10 MHz.
System
Service
Adjustments
Содержание N5242A
Страница 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Страница 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Страница 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Страница 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Страница 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Страница 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Страница 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...
Страница 342: ...Index 22 Service Guide N5242 90001 Index...