28
Installation Note N5230-90022
Step 14. Enable Option F40
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A mouse must be connected to the analyzer for this procedure.
Option Enable Procedure
1. On the analyzer’s
System
menu, point to
Service
, and then click
Option Enable
.
2. In the
Select Desired Option
list, click
F40 - 40 GHz Frequency Rang
e.
3. Click
Enable
.
4. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
5. When the installation is complete, click
Exit
.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s Help menu, click About Network Analyzer.
2. Verify that “F40” is listed after “Options:” in the display. Click OK.
NOTE
If Option F40 has not been enabled, perform the
again. If the option is still not enabled, contact Agilent
Technologies. Refer to