4-28
Service Guide N5230-90014
Troubleshooting
PNA Series Microwave Network Analyzers
Measurement System Troubleshooting
N5230A
Verifying the A, B, R1, and R2 Traces (Standard S-Parameter Mode)
NOTE
There is no way to view the frequency offset receiver response (Option 080).
However, some standard S-parameter receiver trace information is helpful in
troubleshooting the frequency offset section of the PNA. It is therefore
recommended that you run this test even if you
only
suspect the frequency
offset section of malfunctioning.
The first step is to verify that the A, B, R1, and R2 traces are present and that they are
approximately level:
• On the
System
menu, point to
Service
,
Utilities
, and then click
Receiver Display
.
• Traces A, B, R1, and R2 are displayed in four separate data windows as shown in the
. Identifying discrepancies of the traces in these
windows can help you to isolate the faulty assembly.
Figure 4-9 Typical Four Channel Display
Содержание N5230A
Страница 10: ...Contents 6 Service Guide N5230 90014 Contents...
Страница 11: ...Service Guide N5230 90014 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5230 90014 2 1 2 General Product Information...
Страница 31: ...Service Guide N5230 90014 3 1 3 Tests and Adjustments...
Страница 71: ...Service Guide N5230 90014 4 1 4 Troubleshooting...
Страница 121: ...Service Guide N5230 90014 5 1 5 Theory of Operation...
Страница 147: ...Service Guide N5230 90014 6 1 6 Replaceable Parts...
Страница 183: ...Service Guide N5230 90014 7 1 7 Repair and Replacement Procedures...
Страница 229: ...Service Guide N5230 90014 8 1 8 General Purpose Maintenance Procedures...
Страница 262: ...Index 14 Service Guide N5230 90014 Index...