Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Troubleshooting Performance Test Problems
1-74
FM Distortion
Replace the A3 RF assembly
Replace the A3 RF assembly
Digital Modulation Quality
(N5162A/82A only)
Replace the A2 Vector module
N/A
DCFM Carrier Offset
Replace the A3 RF assembly
Replace the A3 RF assembly
Amplitude Switching Speed
Replace the A3 RF assembly
Replace the A7 Micro Deck
If this does not correct the problem, replace the
A3 RF assembly
.
Frequency Switching Speed
Replace the A3 RF assembly
Replace the A3 RF assembly
a
.
a
After replacing the A7 Micro Deck or the A3 RF assembly, go to
Chapter 4, "Post-Repair Procedures and Performance Verification"
Performance Test that is Failing...
Action to Perform...
(N5161A/81A/62A/82A)
Action to Perform...
(N5183A)
深圳德标仪器