Agilent N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Troubleshooting
Troubleshooting Assembly Level Problems
1-39
— If the voltages are not present, replace the A9 Interface Board.
— If the voltages are present, proceed to
•
If A9J2-4 is not
∼
12V, measure A9J2-3.
— If J2-3 is
∼
12V, replace the A9 Interface Board.
— If J2-3 is not
∼
12V, go to
Figure 1-14 A9 Interface Board (A9J2-3 and A9J2-4 Test Points)
9.
Refer to
. Using the voltmeter measure A3J6-4.
•
If A3J6-4 is
∼
12V replace the W1 ribbon cable.
•
If A3J6-4 is not
∼
12V go to
A9J2-3
A9J2-4
A9J2-34