Agilent Technologies N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Post-Repair Procedures and Performance Verification
Leveled Output Power Verification (N5161A/62A
1
/81A/82A Only)
4-15
N5162A
1
/82A Output Power Test Procedure
1.
Zero and calibrate the power sensor to the power meter:
2.
Connect the equipment as shown below:
3.
Preset the signal generator: Press
Preset
.
4.
Select the factory-supplied waveform
SINE_TEST_WFM
:
a.
Press
Mode > Dual ARB > Select Waveform
.
b.
Highlight the
SINE_TEST_WFM
waveform.
c.
Press
Select Waveform
.
5.
Turn the arbitrary waveform player on: Press the
ARB
softkey to highlight
On
.
6.
Turn RF on: Press
RF On/Off
so that the RF On/Off LED lights.
7.
Verify modulation is off:
Press
Mod On/Off
so that the Mod On/Off LED turns off.
8.
Set the signal generator frequency to the first value listed in
:
Press
Frequency
>
338
>
MHz
.
1
The N5161A/62A performs the identical internal tests as in the MXG via a LAN connection to a Web Server. The front panel displays all hardkeys, softkeys and test results
as depicted in the illustration. For information on Web-Enabled Control refer to
“Accessing the MXG Web-Enabled Page” on page 1-18
and to the
Programming
Guide.
Table 4-11 Leveled Output Power Relative to
Frequency
(MHz)
Limit
Δ
(dB)
338
0.25dB
425
0.25dB
538
0.25dB
675
0.25dB
850
0.25dB
1075
0.25dB
1350
0.25dB
1700
0.25dB