Agilent Technologies N5161A/62A/81A/82A/83A MXG Signal Generators Service Guide
Post-Repair Procedures and Performance Verification
Running Self Test
4-9
Use the following procedure to run the self test:
1.
Disconnect all unnecessary
1
external cables, including GPIB, LAN, and USB cables.
2.
Preset the signal generator: Press
Preset > Utility > Instrument Info > Self Test
.
The following message appears:
•
The current status of the self-test is: Incomplete. Not all tests have been run.
3.
Press
Run Complete Self Test
.
An activity bar displays on the screen indicating the test progress.
When self-test completes, one of the following messages displays:
•
The current status of the self-test is: Passed
•
The current status of the self-test is: Failure. One or more tests have failed. System
diagnostics indicate this test as the root failure: xxx
If the signal generator fails only one test, the title of the failed test displays. If the signal generator fails more than one test, the test
number of the most significant failure displays.
Self Test Failure
1.
Confirm that all unnecessary
external cables, including GPIB, LAN, and USB cables, are disconnected from the signal generator and
repeat the self-test.
2.
If the self test continues to fail, the signal generator requires service. If you are unable to service the signal generator, send it to an
Agilent service center for repair. Include a detailed description of the failed test(s) and any displayed error messages.
See
“Viewing Test Results” on page 4-9
for information about viewing detailed self test results.
See
“Returning a Signal Generator for Service” on page 1-2
for return instructions.
Viewing Test Results
If a self test fails the summary display indicates the probable cause.
To view the failed self test
1.
Press:
Utility
>
Instrument Info
>
Self Test
>
View Test Info
2.
Use the arrow keys to highlight the desired test and press
View Details
.
NOTE
This information is
supplemental
. Please disregard unless Agilent
specifically requests
the information.
1
The N5161A/62A require a LAN or USB connection to access the Self Test softkeys via the Web-Enabled MXG. Refer to
“Accessing the MXG Web-Enabled Page” on