Operation
PRBS Generator 10 Gb/s User Guide
25
Connector Description DC Level
DC Level
Z0
Notes
M1 – M3
Mark/space
select
N/A
N/A
N/A
Selected by closure
to lower pin. Refer
to Figure 6.
Reset
Reset PRBS N/A
N/A
N/A
Selected by closure
to lower pin. Refer
to Figure 6.
The turning of the cable or connector while connecting the coupling nut of
the connector turns the center contacts while they are engaging. This wears
the plating, thereby decreasing the signal integrity and operating life of the
connectors. Hold the body of the connector while using a torque wrench to
loosen the coupling nut.
4.5
System Verification
Figure 7 shows a configuration used to verify the performance of the PRBS
source with an external clock. The N4970A internal clock can also be used,
with the CLK/16 output triggering a sampling oscilloscope.
This verification confirms the operation of five pattern lengths, mark density,
and waveshape characteristics.
Содержание N4970A
Страница 1: ...Agilent N4970A PRBS Generator 10 Gb s User Guide ...
Страница 6: ......
Страница 8: ...Getting Started 8 PRBS Generator 10 Gb s User Guide ...
Страница 30: ... Copyright Agilent Technologies 2012 Second edition May 2013 Printed in Germany ...