Using the Multiport Test System
How to Characterize Balanced Devices
Chapter 3
61
Single-Ended Performance
The device has the following single-ended S-parameters:
The parameters along a diagonal (from the upper-left corner to the lower-right corner) on the
display are the single-ended matches on each terminal.
• The six forward-transmission path plots are displayed below the diagonal.
• The six reverse-transmission path plots are displayed above the diagonal.
At low frequencies, the through connection between port 1 and port 2 can be seen from S21
and between port 3 and port 4 from S43. Coupling between the pairs is evident from S41 and
S32 increasing with frequency at the low end of the band while S21 and S43 are decreasing.
Near-end coupling between the input ports (S31) and the output ports (S42) is very low.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...