Maintenance and Troubleshooting
Care of Test Cable Assemblies
Chapter 6
156
Depending on the connector, over-torque can cause damage to connectors in a variety of ways:
• mushroomed outer interface shells
• mushroomed pin shoulders
• recessed or protruding pins
• recessed or protruding dielectrics
• bent pins
• chipped plating
• coupling nut retaining ring damage
• damage to coupling threads
Cable Handling
The test cables have a 1 inch minimum bend radius. Exceeding this radius can result in poor
measurements and poor repeatability. Be alert to tight bends where they are not necessarily
obvious — like at the end of connector strain relief tubing.
Swept 90
°
adapters may be used (typically, at the test set front panel) to minimize cable
bending. However, if the adapters are used, they must be in place during the calibration.
Cables are often stored in a coiled configuration. Coiled cable “set” (the tendency to stay
coiled) can occur if the cables are left coiled. Use large coil diameters (one or two feet) to
minimize cable set. Unroll coiled cables prior to use — never just pull out the loops.
Avoid pinching, crushing, or dropping objects on cable assemblies. Dragging a cable over a
sharp edge will tend to flatten one side, and it is highly likely that the minimum bend radius
will be exceeded.
Содержание N4413A
Страница 4: ...iv ...
Страница 11: ...1 1 About the Multiport Test System ...
Страница 18: ...About the Multiport Test System Available Options and Accessories Chapter 1 8 ...
Страница 19: ...9 2 User Interface ...
Страница 27: ...17 3 Using the Multiport Test System ...
Страница 89: ...79 4 Performing Time Domain Measurements ...
Страница 109: ...99 5 Menu Reference ...
Страница 163: ...153 6 Maintenance and Troubleshooting ...
Страница 172: ...Maintenance and Troubleshooting Contacting Agilent Chapter 6 162 ...
Страница 173: ...163 7 Instrument Information ...
Страница 208: ...Instrument Information Specifications and Characteristics Chapter 7 198 ...
Страница 209: ...199 8 Safety and Regulatory Information ...
Страница 217: ...Safety and Regulatory Information Regulatory Information Chapter 8 207 Declaration of Conformity ...
Страница 218: ...Safety and Regulatory Information Regulatory Information Chapter 8 208 ...
Страница 219: ...209 A Other Technical Procedures ...
Страница 224: ...Other Technical Procedures IF Gain Adjustment Appendix A 214 ...