![Agilent Technologies N4000A Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/agilent-technologies/n4000a/n4000a_operating-and-service-manual_2868361030.webp)
20
Chapter 1
General Information
Operation
Step 9. If this value has not changed, warm the SNS by wrapping your hand
around its bulkhead as shown in Figure 1-11 for 20 seconds. This gives
the SNS thermometer time to change its ambient value. Then press the
User Tcold From SNS
menu key again. The new ambient temperature
value is displayed.
Figure 1-11
Operator’s Check for T
cold
Upload From SNS
Step 10. If the value does not change, return the SNS to Agilent Technologies for
repair.