Voltage Derating

The following graph shows the voltage versus frequency derating curve for the Agilent N2771A High 
Voltage Probe.

Voltage Versus Frequency Derating Curve

Frequency Compensation Adjustment

The N2771A High Voltage Probe has a Frequency Compensation adjustment that compensates for 
the input capacitance of the oscilloscope being used with the probe. 

Use a square-wave generator set for approximately 100 Hz output.  Proceed with the following steps 
for frequency compensation.

1

Connect the probe to the oscilloscope.

2

Connect probe tip to square-wave generator.

3

Adjust the square wave generator for approximately 10 volts amplitude.

4

Adjust the oscilloscope time base for 20 usec/div.

5

Use a trimmer tool to adjust the compensation capacitor for the flattest pulse top.  See the figure 
below for the location of the adjustment.

!

Frequency Compensation Adjustment