8
Installation
User Guide
89
Install the fan module
Decide on which slot(s) to use for the position of your module(s). First consider inlet and
detector locations in the GC oven. If using a detector in the top, front position, then using a
lower slot in the door for the module will provide better clearance for the union and column
ends.
1
Remove the slot cover using a T-10 Torx driver to remove the two screws. See
•
For a 5-inch column module, remove
both
slot covers in the top or bottom row.
Figure 44.
Slot covers. Two top slot covers shown above installed 5-inch column module.
2
Remove the insulation from the slot opening by gently pressing it out from the inside of the
LTM oven door.
3
Seal the cover plate, insulation, and screws in a plastic bag or other sealed container.
Retain in case you reconfigure the LTM system and need to close this slot.
Slot covers