Safety
Physical Hazard Safety
8
K6460 Triple Quad MS Safety Guide
Temperature safety
WA R N I N G
The use of incorrect or makeshift fuses or the short-circuiting of fuse holders
creates a shock hazard for the operator and can damage the instrument. Replace
fuses only with fuses of identical current rating and type. Refer to the
Maintenance
Guide
for more information.
WA R N I N G
Excessive fluctuations in the line voltage can create a shock hazard and can
damage the instrument. This equipment must be installed in a Category II
environment as defined in IEC 664.
For all instruments, the supply voltage must not fluctuate more that ±10%.
For any instrument with an Agilent Jet Stream source, the line voltage must not
fluctuate more that +10/-5% from the rated voltage.
WA R N I N G
Some parts of the K6460 mass spectrometer operate at temperatures high enough
to cause serious burns. Always cool heated parts of the LC/MS down to room
temperature before you work on them.
To cool the instrument faster, set the heated zone to room temperature. Turn the
zone off after it has reached the set point.
If you must do maintenance work on a hot part, use a wrench and wear gloves.