Safety

Physical Hazard Safety

8

K6460 Triple Quad MS Safety Guide

Temperature safety

WA R N I N G

The use of incorrect or makeshift fuses or the short-circuiting of fuse holders 
creates a shock hazard for the operator and can damage the instrument. Replace 
fuses only with fuses of identical current rating and type. Refer to the 

Maintenance 

Guide

 for more information.

WA R N I N G

Excessive fluctuations in the line voltage can create a shock hazard and can 
damage the instrument. This equipment must be installed in a Category II 
environment as defined in IEC 664.

For all instruments, the supply voltage must not fluctuate more that ±10%.

For any instrument with an Agilent Jet Stream source, the line voltage must not 
fluctuate more that +10/-5% from the rated voltage.

WA R N I N G

Some parts of the K6460 mass spectrometer operate at temperatures high enough 
to cause serious burns. Always cool heated parts of the LC/MS down to room 
temperature before you work on them.

To cool the instrument faster, set the heated zone to room temperature. Turn the 
zone off after it has reached the set point. 

If you must do maintenance work on a hot part, use a wrench and wear gloves.

Содержание K6460

Страница 1: ...formity and Standards 12 Safety and Regulatory Certifications 12 Electromagnetic Compatibility 13 Environmental Conditions 14 Sound Emission Declaration 14 This guide contains information about safety...

Страница 2: ...s The following table describes the electrical symbols that can be displayed on the K6460 mass spectrometer Table 1 Electrical symbols Symbol Description Indicates a protective earth terminal Indique...

Страница 3: ...operator and to protect the apparatus against damage L appareil est marqu avec ce symbole lorsque l utilisateur doit se r f rer au manuel d instruction afin de prot ger contre les atteintes l op rateu...

Страница 4: ...l environnement constitue un danger certain Do not discard this electrical electronic product in domestic household waste Ne jetez pas ce produit lectrique lectronique avec les ordures m nag res Disco...

Страница 5: ...Read and understood all material safety data sheets MSDS for chemicals handled WARNING PHYSICAL INJURY HAZARD Use the K6460 mass spectrometer only as specified in the user instructions Using the syst...

Страница 6: ...r cord and the AC power supply In the AC power supply In the wiring from the AC power supply to the power switch WARNING Never remove a cover unless specifically instructed to do so With the power swi...

Страница 7: ...s with adequate current capacity The branch circuit over current protection Circuit Breakers protecting the building receptacles must be rated no more than 20A WARNING Be sure that the LC MS main supp...

Страница 8: ...s equipment must be installed in a Category II environment as defined in IEC 664 For all instruments the supply voltage must not fluctuate more that 10 For any instrument with an Agilent Jet Stream so...

Страница 9: ...es or protective clothing Minimize the inhalation of chemicals Do not leave chemical containers open Use only with adequate ventilation for example fume hood Check regularly for chemical leaks or spil...

Страница 10: ...n a fume hood After emptying a waste container seal it with the cap provided Dispose of the contents of the waste tray and waste bottle in accordance with good laboratory practices and local state pro...

Страница 11: ...r aerosol protection Always follow local state and federal biohazard handling regulation when disposing of biohazardous waste material Handle all infectious samples according to good laboratory proced...

Страница 12: ...gies the protections provided by the instrument may be impaired Safety and Regulatory Certifications The Agilent 6460 Triple Quad Mass Spectrometer Model K6460 conforms to applicable editions and vers...

Страница 13: ...off and on try one or more of the following measures 1 Relocate the susceptible equipment 2 Move the mass spectrometer away from the susceptible equipment 3 Plug the devices into different electrical...

Страница 14: ...kpegel Lp 70 dB Environmental Conditions Equipment Class Class I Laboratory Equipment Pollution Degree 2 Installation Category II Environment Indoor Use Altitude Not to exceed 3000 m Electrical supply...

Страница 15: ...This page intentionally left blank...

Страница 16: ...ry information for your Agilent 6460 Triple Quad Mass Spectrometer Model K6460 system For In Vitro Diagnostic Use The K6460 mass spectrometer is intended to be used to identify inorganic or organic co...

Отзывы: