Relative Time and Voltage
In the table at the bottom of the
Parameters
tab you can change
measurement points. Enter the
Relative Time
and
Voltage
according to
the above settings.
Pass/Fail Tab
The
Pass/Fail
tab of the
Properties
dialog box allows you to specify the
criteria to decide whether the DUT passes or fails the test. You can
change pass/fail criteria without rerunning a test. The software only
uses the criteria to rate the results of a measurement.
N O T E
The pass/fail criteria do not control measurement execution. The measurement run
will be completed even if the measurement fails one or more of the criteria.
The Fast Eye Mask measurement has only one pass/fail condition: The
bit error rate. If desired, enable the
BER Threshold
and set a suitable
threshold.
N O T E
The BER Threshold will usually be 0, since you want the measurement to fail if a
single bit was received in error.
The pass/fail threshold applies to all measurement points. An
icon indicates all measurements where the bit error rate is higher than
this threshold, as shown in the figure below:
View Tab
The following parameters can be set on the
View
tab of the
Properties
dialog box:
You can analyze for:
•
All Errors
To display all errors.
5
Advanced Analysis
248
Agilent J-BERT N4903 High-Performance Serial BERT
Analyze
Содержание J-BERT N4903
Страница 1: ...S Agilent J BERT N4903 High Performance Serial BERT User Guide s Agilent Technologies...
Страница 68: ...2 Setting up Patterns 68 Agilent J BERT N4903 High Performance Serial BERT...
Страница 158: ...4 Setting up the Error Detector 158 Agilent J BERT N4903 High Performance Serial BERT...
Страница 314: ...6 Evaluating Results 314 Agilent J BERT N4903 High Performance Serial BERT...
Страница 374: ...7 Jitter Tolerance Tests 374 Agilent J BERT N4903 High Performance Serial BERT...
Страница 394: ...8 Solving Problems 394 Agilent J BERT N4903 High Performance Serial BERT...
Страница 434: ...Index 434 Agilent J BERT N4903 High Performance Serial BERT...