Agilent InfinityLab ELSD User Manual
43
4
Using the Module
Instrument Controls
Instrument Controls
The ELS Detector can be used as a standalone detector via the front keypad and
screen, as shown in
on page 43 or via PC control using software (e.g.
OpenLab CDS ChemStation).
Display Screen
The graphical interface on the front of the instrument displays the current
method, status, evaporator temperature, nebulizer temperature, gas flow, and
output of the instrument. Operating parameters can be altered via the interactive
menu bar at the bottom of the display.
Figure 7
ELSD display screen
Keypad
The four arrows on the front of the instrument are used to navigate within the
interactive menu bar. The
AZ/Stop
key has a dual function; it can be used to auto
zero the ELSD at any time, unless a timetable is running. If the
AZ/Stop
key is
pressed during an active timetable, the timetable will stop running and the ELSD
will revert to
STANDBY
mode.