Agilent InfinityLab ELSD User Manual
29
3
LAN Configuration
TCP/IP Parameter Configuration
TCP/IP Parameter Configuration
To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. The parameters are:
•
IP address
•
Subnet Mask
•
Default Gateway
The TCP/IP parameters can be configured by the following methods:
•
by manually setting the parameters via the front panel of the detector,
•
by manually setting the parameters through the G7102A ELSD web-interface,
•
by manually setting the parameters through a Telnet session.
The module can be configured, via the front panel, to obtain an IP address
automatically or to use a static IP address (see
configured to obtain its IP address automatically, the module will request an IP
address from the DHCP server after every power cycle.
NOTE
If the detector is assigned a temporary IP address (e.g. DHCP server), then
communication with the PC may be prevented following a power cycle of the
instrument.
Table 8
Default STATIC/IP parameters
IP address
192.168.254.27
Subnet Mask
255.255.255.0
Default Gateway
192.168.254.1