IDP-15 Dry Scroll Vacuum Pump
25
DRAFT
12
/21/17
Run Currents
Typical full load run currents are listed in Table 3.
Start Current
Table 4 lists the typical IDP-15 maximum starting current and duration.
Table 3 IDP-15 Typical Full Load Run Currents (A)
Frequency
100 V
Nominal
120 V
Nominal
200 V
Nominal
240 V Nominal
100 V
120 V
200 V
230 V
240 V
60 Hz
8.05
7.09
4.01
3.60
3.54
50 Hz
7.80
7.93
3.90
3.90
3.94
Table 4 Typical IDP-15 Maximum Starting Current and Duration
Mains
Voltage
50 Hz
60 Hz
Current (A)
Time (ms)
Current (A)
Time (ms)
100 V
22.6
< 1 sec.
23.5
< 1 sec.
120 V
25
< 1 sec.
21.2
< 1 sec.
220 V
11.9
< 1 sec.
14.2
< 1 sec.
230 V
11.7
< 1 sec.
13.9
< 1 sec.