+3$¶$3HUIRUPDQFH7HVW
HP 81130A/’31A Performance Test
7HVW)UHTXHQF\
7HVW6SHFLILFDWLRQV
Range 170 kHz to 400 MHz
Resolution 4 digits, best case 2 ps
Accuracy
+
0.01%
(TXLSPHQW1HHGHG
Counter HP 53132A
Cable, 50
Ω
, coaxial, BNC; SMA/BNC Adaptor
127(
The HP 53132A counter is used in frequency mode to meet the
MIL CAL A uncertainty requirements for TAR (Test Accuracy
Ratio) > 4:1.
3URFHGXUH
Connect the HP 81130A to the counter as follows:
Connecting HP 81130A to the Counter
4. Set up the HP 81130A as described in "Initial Setup of the HP
81130A"
+3$81'(57(67
TRIG OUT
+3$&RXQWHU
CHANNEL 1
2
3
BNC Cable
SMA/BNC
Adaptor