Interface Capabilities
The baud rate, stop bits, parity, protocol, and data bits must be configured
exactly the same for both the controller and the logic analysis system to
properly communicate over the RS-232C bus. The RS-232C interface
capabilities of the HP 16500B Logic Analysis System are listed below:
•
Baud Rate: 110, 300, 600, 1200, 2400, 4800, 9600, or 19.2k
•
Stop Bits: 1, 1.5, or 2
•
Parity: None, Odd, or Even
•
Protocol: None or XON/XOFF
•
Data Bits: 8
Protocol
NONE
With a three-wire interface, selecting NONE for the protocol
does not allow the sending or receiving device to control data flow. No
control over the data flow increases the possibility of missing data or
transferring incomplete data.
With an extended hardwire interface, selecting NONE allows a hardware
handshake to occur. With hardware handshake, the hardware signals control
data flow.
XON/XOFF
XON/XOFF stands for Transmit On/Transmit Off. With this
mode, the receiver (controller or logic analysis system) controls
data flow, and, can request that the sender (logic analysis system or
controller) stop data flow. By sending XOFF (ASCII 19) over its transmit
data line, the receiver requests that the sender disables data
transmission. A subsequent XON (ASCII 17) allows the sending device
to resume data transmission.
Data Bits
Data bits are the number of bits sent and received per character that
represent the binary code of that character. Characters consist of either 7 or
8 bits, depending on the application. The HP 16500B Logic Analysis
System supports 8 bit only.
8 Bit Mode
Information is usually stored in bytes (8 bits at a time).
With 8-bit mode, you can send and receive data just as it is stored,
without the need to convert the data.
Programming Over RS-232C
Interface Capabilities
3–10
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Содержание HP 16500B
Страница 13: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 37: ...1 24 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 97: ...7 6 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 99: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 100: ...8 Common Commands 8 1 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 119: ...8 20 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 120: ...9 Mainframe Commands 9 1 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...
Страница 146: ...10 SYSTem Subsystem 10 1 Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com ...