![Agilent Technologies G7035A Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/agilent-technologies/g7035a/g7035a_operation-manual_2868127099.webp)
Operating in Chemical Ionization (CI) Mode
4
5977 Series MSD Operation Manual
99
To Set Up the Software for CI Operation
Procedure
1
From the
Tune and Vacuum Control
view, select
Load Tune Parameters
from the
File
menu and load the tune file
PCICH4.U
.
2
If CI autotune has never been run for this tune file, the software will
prompt you through a series of dialog boxes.
Accept the default values
unless you have a very good reason for changing anything.
The tune values have a dramatic effect on MSD performance. Always start
with the default values when first setting up for CI, and then make
adjustments for your specific application. See
for default values
for the Tune Control Limits box. These limits are used by Autotune only.
They should
not
be confused with the parameters set in Edit MS
Parameters or with those appearing on the tune report.
Notes for
:
• N/A
Not available. There are no PFDTD ions formed in PCI with any
reagent gas but methane, hence, CI autotune is not available with these
configurations.
C A U T I O N
Always verify GC/MS performance in EI before switching to CI operation.
Table 17
Default Tune Control Limits, used by CI autotune only
Reagent gas
Methane
Isobutane
Ammonia
Ion polarity
Positive
Negative
Positive
Negative
Positive
Negative
Abundance target
1x10
6
1x10
6
N/A
1x10
6
N/A
1x10
6
Peakwidth target
0.6
0.6
N/A
0.6
N/A
0.6
Maximum repeller
4
4
N/A
4
N/A
4
Maximum emission
current, µA
240
50
N/A
50
N/A
50
Max electron energy, eV
240
240
N/A
240
N/A
240
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...