38
5977 Series MSD Operation Manual
2
Installing GC Columns
To Install a Capillary Column in the GC/MS Interface
This procedure is for the installation of a capillary column directly to the
transfer line with a column nut. If you are using the Agilent capillary flow
technology Quick Swap accessory, or a Purged Ultimate Union (PUU) see
Prepare the Column Ends for a CFT Fitting”
Agilent 7890 Series GC
Materials needed
•
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
•
Ferrules
•
0.3 mm id, for 0.10 mm id columns (5062-3507)
•
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
•
0.5 mm id, for 0.32 mm id columns (5062-3506)
•
0.8 mm id, for 0.53 mm id columns (5062-3512)
•
Flashlight
•
Magnifying loupe
•
Gloves, clean
•
Large (8650-0030)
•
Small (8650-0029)
•
Interface column nut (05988-20066)
•
Safety glasses
•
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)
•
Column measuring tool
C A U T I O N
Always wear clean gloves while handling any parts that go inside the GC or the
analyzer chambers.
Содержание G7035A
Страница 1: ...Agilent Technologies Agilent 5977 Series MSD Operation Manual ...
Страница 30: ...30 5977 Series MSD Operation Manual 1 Introduction ...
Страница 88: ...88 5977 Series MSD Operation Manual 3 Operating in Electron Ionization EI Mode ...
Страница 112: ...112 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 18 PCI autotune report ...
Страница 114: ...114 5977 Series MSD Operation Manual 4 Operating in Chemical Ionization CI Mode Figure 19 NCI autotune ...
Страница 170: ...170 5977 Series MSD Operation Manual 6 CI Maintenance ...
Страница 171: ......
Страница 172: ... G3870 90003 G3870 90003 Agilent Technologies Agilent Technologies Inc Printed in USA February 2013 ...