CI Maintenance
6
5977 Series MSD Operation Manual
167
To Install the CI Ion Source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Pliers, long-nose (8710-1094)
Procedure
1
Slide the ion source into the source radiator.
2
Install and hand tighten the source thumbscrews. Do not overtighten the
thumbscrews.
3
Connect the ion source wires as shown in
“To Close the Analyzer Chamber”
4
Close the analyzer chamber. See
“To Close the Analyzer Chamber”
Source feedthrough board
Ion source
Thumbscrews
Source heater and temperature
sensor wires
Source radiator