80
1260 SFC Binary Pump
6
Error Information
Module Error Messages
Missing Pressure Reading
Error ID: 2054
The pressure readings read by the pump ADC (analog-digital converter) are
missing.
The ADC reads the pressure signal of from the damper every 1ms. If the
readings are missing for longer than 10 s, the error message is generated.
Wrong Pump Configuration
Error ID: 2060
At switch-on, the pump has recognized a new pump configuration.
The binary pump is assigned its configuration at the factory. If the active inlet
valve and pump encoder of channel B are disconnected, and the binary pump
is rebooted, the error message is generated.
Probable cause
Suggested actions
1
Damper disconnected.
Please contact your Agilent service
representative.
2
Defective damper.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Active-inlet valve and pump encoder of
channel B disconnected.
Please contact your Agilent service
representative.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...