26
1260 SFC Binary Pump
2
Site Requirements and Specifications
Site Requirements
Bench Space
The module dimensions and weight (see
on page 27) allow you to place
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm
(3.1 inches) in the rear for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
The module should be operated in a horizontal position.
Condensation
WA R N I N G
Power cords
Solvents may damage electrical cables.
➔
Prevent electrical cables from getting in contact with solvents.
➔
Exchange electrical cables after contact with solvents.
N O T E
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench
rack. This option helps to save bench space as all modules can be placed into one single
stack. It also allows to easily relocate the instrument to another Lab.
C A U T I O N
Condensation within the module
Condensation can damage the system electronics.
➔
Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
➔
If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.
Содержание G4782A
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series 1260 Infinity II SFC Binary Pump User Manual...
Страница 8: ...8 1260 SFC Binary Pump Contents...
Страница 22: ...22 1260 SFC Binary Pump 1 Introduction Leak and Waste Handling...
Страница 30: ...30 1260 SFC Binary Pump 2 Site Requirements and Specifications Performance Specifications...
Страница 66: ...66 1260 SFC Binary Pump 5 Troubleshooting and Diagnostics Agilent Lab Advisor Software...
Страница 140: ...140 1260 SFC Binary Pump 9 Parts and Materials for Maintenance Accessory Kit...
Страница 172: ...172 1260 SFC Binary Pump 11 Hardware Information Instrument Layout...