8
Install the Capillaries
Install the Valve Head and Connect Capillaries
Install the Capillaries
3
Install all required capillary connections to the valve.
4
Push the valve head until it snaps in and stays in the rear
position.
5
Power on or power-cycle your module, so the valve head gets recognized during module initialization.
Damage to the rotor seal
Instant pressure release within the valve will lead to water jet effects that can harm internal
parts of the valve. This pressure release typically happens if the valve gets switched under high
pressure over unused or open channels.
➔
Block all unused channels properly with the M4 blank nut.
C A U T I O N
N O T E
To minimize valve movement over open connections it is recommended to plumb the column
connected channels in one row.
e.g.:
•
channel 1 – column 1
•
channel 2 – column 2
•
channel 3 – column 3
•
channel 4 – column 4
•
channel 5 – blocked
•
channel 6 – blocked
•
channel 7 – waste
•
channel 8 – bypass
N O T E
The blank nuts are only required for the ports on the inner circle that connect the valve with the
column inlet.