20
Installation Instructions
Agilent G3440-68011 Micro Electron Capture (µECD) 3rd Detector
Figure 26
Ferrules and nut placement
7
Install the
μ
ECD vent tube onto the
μ
ECD module and route the
μ
ECD gas
tubing to the detector body as shown in
.
8
Route the
μ
ECD vent tube out the back of the 3rd detector box assembly.
Figure 27
Install µECD vent tube
9
Route the gas tubing so that it does not interfere with any other parts inside
the box frame. Use gentle bends with a large radius. Do not kink or pinch
the tubing.
10
Secure the
μ
ECD serial tag to the 3rd detector box assembly using a tie
wrap as shown in
.
Tubing
Front ferrule
Back ferrule
1/8-inch Swagelok nut