9
Agilent Technologies
Agilent G3388A Leak Detector
Operation Manual
2
Operating Procedure
1
The detector’s batteries are not fully charged when they are
shipped from the factory. Recharge the batteries before using
the detector.
2
Wait for 30 minutes after recharging for the batteries’
temperature to stabilize.
3
Turn the power switch to the ON position and allow the
detector to warm up for approximately 25 seconds. If the
buzzer sounds during this time, press the
ZERO/ENTER
key to
stop it.
4
When the LCD shows the following display (Standard
display), press the
ZERO/ENTER
key. Now the detector is ready
to use. “Range_L” may be displayed if this was the last
setting.
Indicating the countdown of the warm-up time by
seconds.
Press the ZERO/ENTER key to skip the warm-up step.
Warming-up display
b. Battery display
a. Range indicator
c. Leak indicator
Standard display
a
Range indicator
Range_H:
High sensitivity
Range_L:
Low sensitivity