To adjust or replace the nebulizer needle
10
Nebulizer Adjustment Fixture User Guide
If the
is pushed in too far, when the union is installed and the
is compressed, the HPLC flow is restricted or blocked, which
results in poor or no spray from the nebulizer.
h
Reinstall the locknut and the union. Hand-tighten the union.
i
Hold the
steady with a
, depending on your nebulizer holder style.
j
Tighten the union one-half to three quarters of a turn, or until an audible
squeak is heard.
The ferrule is now compressed.
8
Use a
¼-inch x 5/16-inch open-end wrench
to adjust the top of the
the proper distance from the end of the nozzle. See
.
9
Tighten the lock nut and verify adjustment while you view the nebulizer tip
through the magnifier.
See
for images of correctly adjusted nebulizer needles.
Table 2
Recommended Needle adjustment Distance
Nebulizer
Distance
0.003 inch
0.000 (flush)
0.003 inch
Tested APCI Nebulizer Assembly
0.000 (flush)