Introduction to Headspace and to the G1888 Network Headspace Sampler
End of cycle and vial recovery
The sample vial lowers into the carousel. The carousel rotates
and the lift returns the vial to the tray. The vent valve opens to
purge the vent line. The sampler returns to the standby
configuration. See
. The default vent time for S2 is
30 seconds. To adjust the vent time, see
.
Figure 11
Purge. S1 is open. S2 is open for 30 seconds.
Carrier gas flow
Mass
controller
Pressure
transducer
Vent
Sample
loop
To GC
Auxiliary gas flow
for vial pressurization
Pressure
regulator
Pressure
transducer
Flow restrictor
S1
S2
Needle
Headspace vial
Open
Closed
Carrier gas
Auxiliary gas
1 2
3
4
5
6
flow
Toggle
switch
Toggle
switch
medium_standard.book Page 40 Tuesday, February 17, 2004 10:14 AM
Содержание G1888
Страница 12: ...12 User Information medium_standard book Page 12 Tuesday February 17 2004 10 14 AM...
Страница 24: ...24 User Information Introduction medium_standard book Page 24 Tuesday February 17 2004 10 14 AM...
Страница 118: ...118 User Information Developing Methods medium_standard book Page 118 Tuesday February 17 2004 10 14 AM...
Страница 162: ...162 User Information Maintaining the Instrument medium_standard book Page 162 Tuesday February 17 2004 10 14 AM...
Страница 190: ...190 User Information Index medium_standard book Page 190 Tuesday February 17 2004 10 14 AM...
Страница 191: ...medium_standard book Page 191 Tuesday February 17 2004 10 14 AM...